<7. HYDROGEN PURITY / PARAMETER SETTING>

7-17

IM 11T3E1-01E

(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to “2”.
Configurable range: -10.0 to 110.0 (percent of analog output)

[ENT]

03

*CODE

*SERVC

X

*H_HLD

XXX.X

*PR.SET

[ENT]

[ENT]

[YES]

F0710.ai

*

1

*

2

*

3

*

4

*

5

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ] and [ENT] keys.

*4: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].

Figure 7.10 Process of Setting the Hold of Hydrogen Purity Range

[Note]
Relationship of CODE 01 to 03 with Output Hold Function Applied to Parameter [*CAL.DT] at

Setting level.

Hydrogen purity/Replacement (Substituents-concentration) Meters Mode

Priority

Status

Hydrogen purity Range is

Selected for Output

Hydrogen purity Range is

Not Selected for Output

1

Error

The output value selected in CODE03 The output value selected in CODE 02

2

Calibration

The output value selected in CODE03 The output value selected for data item

[*C_HLD] under parameter [*CAL.DT]

3

Maintenance

The output value selected in CODE03 The output valve selected in CODE 01

4

Switch to replacement

meter mode

The output value selected in CODE03 The normal output condition applies

Example: If you select parameter [*H_A] for output 1 and parameter [*H_C_A] for output 2, then

the resulting status-by-status output values to be held are summarized as follows:

Output 1

Output 2

During error status

As per CODE 03

As per CODE02

During calibration status As per CODE 03

As per data item [*C_HLD ]

under parameter [*CAL.DT]