Chapter 5 To Trigger the Oscilloscope

5-26

MSO1000Z/DS1000Z User’s Guide

Edge B:

Press EdgeB to select the trigger edge type of signal source B in delay trigger. It can

be set to the rising edge or falling edge.

Delay Type:

Press DelayType to set the time limit condition of delay trigger.


>: trigger when the time difference (

△T) between the specified edges of source

A and source B is greater than the preset time lower limit. Press Lower Limit to

set the delay time lower limit in delay trigger and the range is from 8.00 ns to 10

s. For the setting method, please refer to the introduction in "Parameter

Setting Method".

<: trigger when the time difference (

△T) between the specified edges of source

A and source B is lower than the preset time upper limit. Press Upper Limit to

set the delay time upper limit in delay trigger and the range is from 16.0 ns to

10.0 s. For the setting method, please refer to the introduction in "Parameter

Setting Method".

<>: trigger when the time difference (

△T) between the specified edges of

source A and source B is greater than the lower limit of the preset time and

lower than the upper limit of the preset time. Press Upper Limit to set the

upper limit of the delay time in delay trigger and the range is from 16.0 ns to

10.0 s. Press Lower Limit to set the lower limit of the delay time in delay

trigger and the range is from 8.00 ns to 9.99 s. For the setting method, please

refer to the introduction in "Parameter Setting Method".

Note:

The time lower limit must be lower than the time upper limit.

><: trigger when the time difference (

△T) between the specified edges of

source A and source B is lower than the lower limit of the preset time or greater

than the upper limit of the preset time. Press Upper Limit to set the upper limit

of the delay time in delay trigger and the range is from 16.0 ns to 10.0 s. Press

Lower Limit to set the lower limit of the delay time in delay trigger and the

range is from 8.00 ns to 9.99 s. For the setting method, please refer to the

introduction in "Parameter Setting Method".

Note:

The time lower limit must be lower than the time upper limit.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to "

Trigger Mode". The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection)

under this trigger type.

Trigger Level:

The trigger level of each source needs to be set independently. For example, set the

trigger level of source A. Press SourceA to select the desired source, and then use

Advertising