Chapter 6 MATH and Measurement

6-10

MSO1000Z/DS1000Z User’s Guide

"AND" Operation

Perform logic "AND" operation on the waveform voltage values of the specified

sources point by point and display the results. In operation, when the voltage value

of the source channel is greater than the threshold of the corresponding channel, it is

regarded as logic "1"; otherwise logic "0".

The results of logic AND operation of two binary bits are as follows.

Table 6-2 Logic "AND" Operation

A

B

A&&B

0

0

0

0

1

0

1

0

0

1

1

1


Press MATHMathOperator to select "A&&B":
 Press Operation to enable or disable the "AND" operation function.
 Press SourceA and SourceB to select the desired channels (CH1-CH4 or

D0-D15).

If source A selects any channel in CH1-CH4, press Thre.A and use

set

the threshold of source A in logic operation. If source A selects any channel

in D0-D15, Thre.A will be hidden automatically.

If source B selects any channel in CH1-CH4, press Thre.B and use

set

the threshold of source B in logic operation. If source B selects any channel

in D0-D15, Thre.B will be hidden automatically.

 Press Offset and use

to adjust the vertical position of the operation results.

 Press Scale and use

to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the

optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or

disable waveform invert, etc.

Press Start and use

to set the start point of the operation results.

Press End and use

to set the end point of the operation results.

Press Invert to enable or disable the inverted display function of the

waveform.

Press Sens. and use

to set the sensitivity of the digital signal converted

from the analog signal on the source. The setting range is from 0 Div to

0.96 Div.

Press Auto Scale to enable or disable the auto scale function. When auto

scale is enabled, the instrument will adjust the vertical scale of the

operation results to the optimal value according to the current

configuration.