To output arbitrary waveform, To output arbitrary waveform -10 – RIGOL MSO/DS1000Z Series User Manual

Page 206

Advertising
background image

RIGOL

Chapter 13 Signal Source

13-10

MSO1000Z/DS1000Z User’s Guide

8. Modulation

Press Src1Mod to enter the modulation setting interface. For detailed

introduction of the modulation function, please refer to "Modulation".

Note:

When "Pulse", "DC" or "Noise" is selected, the modulation menu will be

hidden automatically.

9. Set the impedance

Press Impedance to set the output impedance of the signal generator. It can be

set to "HighZ" or "50Ω".

To Output Arbitrary Waveform

MSO1000Z/DS1000Z allows users to define their own arbitrary waveform and save it

in the internal or external memory. At most 10 arbitrary waveforms can be stored in

the internal memory. The user-defined waveform can contain 1 to 16384 data points

(namely 1 pts to 16 kpts).

1. Source 1 output

Press Output to turn on or off the signal output.

2. Set the frequency

Press Frequency to set the frequency of the signal. For the setting method,

please refer to the introduction in "

Parameter Setting Method". The available

range is from 100 mHz to 10 MHz.

3. Set the amplitude

Press Amplitude to set the amplitude of the signal. For the setting method,

please refer to the introduction in "

Parameter Setting Method". When the

impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the

impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.

4. Set the DC offset voltage

Press Offset to set the DC offset of the signal. For the setting method, please

refer to the introduction in "Parameter Setting Method". When the

impedance is HighZ, the range is from (-2.5 V + the current amplitude/2) to (2.5

V - the current amplitude/2); when the impedance is set to 50 Ω, the range is

from (-1.25 V + the current amplitude/2) to (1.25 V - the current amplitude/2).

5. Set the start phase

Press StartPhase to set the start phase of the signal. For the setting method,

please refer to the introduction in "

Parameter Setting Method". The range is

from 0° to 360°.