Auto measurement, Quick measurement after auto, Auto measurement -22 – RIGOL MSO/DS1000Z Series User Manual

Page 136: Quick measurement after auto -22

Advertising
background image

RIGOL

Chapter 6 MATH and Measurement

6-22

MSO1000Z/DS1000Z User’s Guide

Auto Measurement

MSO1000Z/DS1000Z provides auto measurements of 33 waveform parameters and

the statistics and analysis of the measurement results. What’s more, you can also

use the frequency counter to realize more precise frequency measurement.

Quick Measurement after AUTO

When the oscilloscope is correctly connected and has detected valid input signal,
press AUTO to enable the waveform auto setting function and open the following

function menu:

Single-period: set the oscilloscope to display single-period signal

automatically. Measure the "Period" and "Frequency" of the

current signal within a single period and display the measurement

results at the bottom of the screen.

Multi-period: set the oscilloscope to display multi-period signal

automatically. Measure the "Period" and "Frequency" of the

current signal within multiple periods and display the

measurement results at the bottom of the screen.

Rise Time: measure the "Rise Time" of the current signal and

display the measurement result at the bottom of the screen.

Fall Time: measure the "Fall Time" of the current signal and

display the measurement result at the bottom of the screen.

Back: return to the menu displayed at the last user setting.

Undo: cancel the automatic settings and restore the parameters

set by users lastly.

Figure 6-2 Function Menu after AUTO

Note:

Waveform auto setting function requires that the frequency of sine is no lower

than 41 Hz; the duty cycle should be greater than 1% and the amplitude must be at

least 20 mVpp for square. Otherwise, the Waveform auto setting function may be

invalid and the quick parameter measurement function displayed in the menu will

also be unavailable.