RIGOL MSO/DS1000Z Series User Manual

Page 209

Advertising
background image

Chapter 13 Signal Source

RIGOL

MSO1000Z/DS1000Z User’s Guide

13-13

OFF: the waveform editor will keep a constant voltage level between two

points and create a ladder-like waveform.

3. Zoom

Press Zoom to turn on or off the zoom function.
ON: the waveform editing window only displays the current point.
OFF: the waveform editing window displays all the initial points.

4. Current Point

Press CurPoint to set the point to be edited. The range is from 1 to the initial

points. For the setting method, please refer to the introduction in "Parameter

Setting Method".

5. Voltage

Press Voltage to set the voltage of the current point. The range is from -2.5 V to

+2.5 V and the default is 0.000 mV. For the setting method, please refer to the

introduction in "Parameter Setting Method".

6. Time

Press Time to set the duration of the current point. This setting is limited by the

times of the previous point and following point. The time of point 1 is fixed at 0 s.

7. Insert

Press Insert to insert a new waveform point between the current point and the

next point. At this time, the initial points increases by one. You can press Insert

continuously to increase the editable point successively.

8. Delete

Press Delete to delete the current point from the waveform and connect the

remaining points using the current interpolation mode.

Note:

Point 1 cannot be deleted.

9. APPly

Press Apply to finish editing the current waveform and output the edited

waveform.

10. Save

Press Save to enter the file store interface. Please refer to the introduction in

Advertising