Chapter 6 MATH and Measurement

RIGOL

MSO1000Z/DS1000Z User’s Guide

6-29

minimum value within the settable range.

3. Delay Measurement Setting

Specify the source1 and source2 in the measurement items "Delay 12 " and
"Delay 1

2 ".

Press MeasureSettingType  "Delay" and then press SourceA and
SourceB
to set the two channel sources (CH1-CH4 or D0-D15) of delay
measurement respectively.

4. Phase Measurement Setting

Specify the source1 and source2 in the measurement items "Phase 12 "
and "Phase 1

2 ".

Press MeasureSettingType  "Phase" and then press SourceA and
SourceB to set the two channel sources (CH1-CH4 or D0-D15) of phase
measurement respectively.

5. Threshold Measurement Setting

Specify the upper limit, lower limit and middle value of the auto measurement of

the analog channel. Measurements of all the time, delay and phase parameters

will be influenced by this setting.
Press MeasureSettingType  "Threshold" and then:
 Press Max and use

to set the upper limit of the measurement.

Reducing the upper limit to the current middle value will automatically
reduce the middle value and lower limit to keep them lower than the upper
limit. The default is 90% and the range available is from 7% to 95%.

 Press Mid and use

to set the middle value of measurement. The middle

value is limited by the settings of the upper limit and lower limit. The default

is 50% and the range available is from 6% to 94%.

 Press Min and use

to set the lower limit of the measurement.

Increasing the lower limit to the current middle value will automatically

increase the middle value and upper limit to keep them higher than the

lower limit. The default is 10% and the range available is from 5% to 93%.

To Clear the Measurement

If you have currently enabled one or more items in the 33 measurement parameters,

you can "Delete" or "Recover" the last five parameters. The measurement results of

measurement items 1 to 5 are displayed from left to right at the bottom of the

screen.

Note:

The last five parameters are determined according to the order in which you

turned them on and they will not change as you delete one or more measurement

items.