Window trigger (option), Window trigger (option) -23 – RIGOL MSO/DS1000Z Series User Manual

Page 97

Advertising
background image

Chapter 5 To Trigger the Oscilloscope

RIGOL

MSO1000Z/DS1000Z User’s Guide

5-23

Window Trigger (Option)

Window trigger provides a high trigger level and a low trigger level. The instrument

triggers when the input signal passes through the high trigger level or the low trigger

level.

Trigger Type:

Press Type, rotate

to select "Window" and press down

. At this point, the

trigger setting information is displayed at the upper right corner of the screen.

For example,

. The trigger type is window trigger; the trigger

source is CH1; the difference between the upper limit of trigger level and the lower

limit of trigger level is 2.00 V.

Source Selection:

Press Source to switch the trigger source and select CH1-CH4. For the details,

please refer to the introduction in "

Trigger Source". The current trigger source is

displayed at the upper right corner of the screen.

Note:

Select channel with signal input as trigger source to obtain stable trigger.

Window Type:

Press WndType to select the kind of edge of the input signal on which the

oscilloscope triggers.

: trigger on the rising edge of the input signal when the voltage level is

greater than the preset high trigger level.

: trigger on the falling edge of the input signal when the voltage level is

lower than the preset low trigger level.

: trigger on any edge of the input signal when the voltage level meets the

preset trigger level.

Trigger Position:

After selecting the window type, press Position to further specify the time point of

trigger by selecting the trigger position.
Enter: trigger when the trigger signal enters the specified trigger level range.
Exit: trigger when the input signal exits the specified trigger level range.
Time: trigger when the accumulated hold time after the trigger signal enters

the specified trigger level range is equal to the window time. When this type is

selected, press Time to set the window time. The range is from 8 ns to 10 s. For

the setting method, please refer to the introduction in "Parameter Setting

Method".

Vertical Window and Trigger Level:

Press Window to select the desired vertical window type and use Trigger

LEVEL to adjust the trigger level. For detailed operation, please refer to the

introduction in "Vertical Window and Trigger Level" in runt trigger.

Advertising