Power-off recall, Self-calibration, Power-off recall -8 – RIGOL MSO/DS1000Z Series User Manual

Page 240: Self-calibration -8

Advertising
background image

RIGOL

Chapter 15 Accessibility Setting

15-8

MSO1000Z/DS1000Z User’s Guide

Power-off Recall

You can set the system configuration to be recalled when the oscilloscope is powered
on again after power-off. Press UtilitySystemPower Set to select "Last" or
"Default".
 Last: return to the setting of the system at last power-off.
 Default: return to the factory setting of the system.

Self-calibration

The self-calibration program can quickly make the oscilloscope reach the best

working state to get the most precise measurement values. You can perform

self-calibration at any time especially when the change of the environment

temperature is up to or more than 5℃. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.


Disconnect all the input channel connections, and then press UtilitySelf-Cal and
the self-calibration interface as shown in the figure below is displayed.

Figure 15-2 Self-calibration Interface

Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the

previous menu.

Note:

Most of the keys are disabled during the self-calibration.