Multiplication, Division, Multiplication -4 – RIGOL MSO/DS1000Z Series User Manual

Page 118: Division -4

Advertising
background image

RIGOL

Chapter 6 MATH and Measurement

6-4

MSO1000Z/DS1000Z User’s Guide

Multiplication

Multiply the waveform voltage values of signal source A and B point by point and

display the results.


Press MATHMathOperator to select "A×B":
 Press Operation to enable or disable the multiplication operation function.
 Press SourceA and SourceB to select the desired channels (CH1, CH2, CH3,

CH4 or fx (please refer to the introduction in "

Fx Operation")).

 Press Offset and use

to adjust the vertical position of the operation results.

 Press Scale and use

to adjust the vertical scale of the operation results.

 Press Scale Reset to adjust the vertical scale of the operation results to the

optimal value according to the current configuration.

 Press Options to set the start and end points of the operation results, enable or

disable waveform invert, etc.

Press Start and use

to set the start point of the operation results.

Press End and use

to set the end point of the operation results.

Press Invert to enable or disable the inverted display function of the

waveform.

Press Auto Scale to enable or disable the auto scale function. When auto

scale is enabled, the instrument will adjust the vertical scale of the

operation results to the optimal value according to the current

configuration.

Press fx Operator, fx A and fx B to set the operator and signal sources of

the inner layer operation of Fx operation (please refer to the introduction in

"Fx Operation").

Note:

Sens. and Smooth are grayed out and disabled. Sens. is only available

when a digital channel is selected as the source. Smooth is only available for

differential operation.

 You can also use HORIZONTAL POSITION and HORIZONTAL

SCALE to adjust the horizontal position and scale of the operation results.

Division

Divide the waveform voltage values of signal source A by that of source B point by

point and display the results. It can be used to analyze the multiple relationships of

waveforms of two channels.

Note:

When the voltage of signal source B is 0V, the division result is invalid and

"NAN" is displayed at the bottom of the screen.


Press MATHMathOperator to select "A/B":
 Press Operation to enable or disable the division operation function.

Advertising