Chapter 18 Specifications

RIGOL

MSO1000Z/DS1000Z User’s Guide

18-1

Chapter 18 Specifications

All the specifications are guaranteed except the parameters marked with "Typical"

and the oscilloscope needs to operate for more than 30 minutes under the specified

operation temperature.

Sample

Sample Mode

Real-time Sample

Real Time

Sample Rate

Analog channe: 1 GSa/s (single-channel), 500 Msa/s

(dual-channel), 250 MSa/s (3/4-channel)

Digital channel: 1 GSa/s (8-channel), 500 MSa/s (16-channel)

Peak Detect

Analog channe: 4 ns

Digital channel: 4 ns

Averaging

After all the channels finish N samples at the same time, N can

be 2, 4, 8, 16, 32, 64, 128, 256, 512 or 1024

High Resolution 12 bit (max)
Interpolation

Sin(x)/x (optional)

Min Detect

Pulse Width

Digital channel: 10 ns

Memory Depth

Analog channel: standard 12M pts (single-channel), 6M pts

(dual-channel), 3M pts (3/4-channel); optional 24 Mpts

(single-channel), 12 Mpts (dual-channel), 6 Mpts (3/4-channel)

Digital channel: standard 12 Mpts (8-channel)/6 Mpts

(16-channel); optional 24 Mpts (8-channel)/12 Mpts

(16-channel)

Input

Number of

Channels

MSO1XX4Z/1XX4Z-S and DS1XX4Z Plus/1XX4Z-S Plus: 4 analog

channels + 16 digital channels

DS1054Z: 4 analog channels

Input Coupling

DC, AC or GND

Input

Impedance

Analog channe: (1 MΩ±1%) || (15 pF±3 pF)

Digital channel: (100 kΩ±1%) || (8 pF±3 pF)

Probe

Attenuation

Coefficient

Analog channe: 0.01X to 1000X, in 1-2-5 step

Maximum Input

Voltage (1MΩ)

Analog Channel:

CAT I 300 Vrms, CAT II 100 Vrms, Transient Overvoltage 1000

Vpk

With RP2200 10:1 probe: CAT II 300 Vrms

Digital channel: CAT I 40Vrms, Transient Overvoltage 800 Vpk