Chapter 6 MATH and Measurement

6-8

MSO1000Z/DS1000Z User’s Guide

Table 6-1 Window Function

Window Characteristics

Waveforms Suitable for

Measurement

Rectangle The best frequency

resolution;

the poorest amplitude

resolution;

similar to the situation

when no window is

multiplied.

Transient or short pulse, the signal levels

before and after the multiplication are

basically the same;

Sine waveforms with the same amplitude

and rather similar frequencies;

Wide band random noise with waveform

spectrum changing relatively slower.

Blackman The best amplitude

resolution;

the poorest frequency

resolution

Single frequency signal, search for

higher order harmonics.

Hanning

Better frequency

resolution and poorer

amplitude resolution

compared to

Rectangle window.

Sine, periodic and narrow band random

noise.

Hamming A little bit better

frequency resolution

than Hanning

Transient or short pulse, the signal levels

before and after the multiplication are

rather different.

Flattop

Accurate

measurement of

signal.

Signal without accurate reference and

require to accurately measurement.

Triangle

Better frequency

resolution.

Narrow band signal with stronger

interference noise.

8. Set the Display Mode

Press View to select "Half" (default) or "Full" display mode.
 Half: the source channel and the FFT operation results are displayed

separately. The time domain and frequency domain signals are displayed

clearly.

 Full: the source channel and the FFT operation results are displayed in the

same window to view the frequency spectrum more clearly and to perform

more precise measurement.

9. Set the Vertical Unit

The vertical unit can be selected as dB/dBm or Vrms. Press Unit to select the

desired unit and the default is dB/dBm. dB/dBm and Vrms use logarithmic mode

and linear mode to display vertical amplitude respectively. If you need to display

the FFT frequency spectrum in a relatively larger dynamic range, dB/dBm is

recommended.

10. Press Scale Reset to adjust the vertical scale of the operation results to the