Chapter 15 Accessibility Setting

RIGOL

MSO1000Z/DS1000Z User’s Guide

15-11

Aux Output

Users can set the type of the signal output from the [Trigger Out] connector at the

rear panel.


Press UtilityAux Out to select the desired output type.

1. TrigOut

After this type is selected, the oscilloscope outputs a signal that can reflect the

current capture rate of the oscilloscope at each trigger. Connect this signal to a

waveform display device, measure the signal frequency and the measurement

result equals the current acquisition rate.

2. PassFail

After this type is selected, the oscilloscope will output a negative pulse from this

connector when a failed waveform is detected in the pass/fail test. The

instrument will output low level continuously if no failure is detected.