Chapter 17 Troubleshooting

RIGOL

MSO1000Z/DS1000Z User’s Guide

17-1

Chapter 17 Troubleshooting

The commonly encountered failures and their solutions are listed below. When you

encounter those problems, please solve them following the corresponding steps. If

the problem remains still, please contact

RIGOL and provide your device

information (Utility SystemSystem Info).

1. The screen is still dark (no display) after pressing the power key:

(1) Check whether the power switch is really on.

(2) Check whether the power is correctly connected.

(3) Check whether the fuse is burned out. If the fuse needs to be changed,

please use the specified fuse.

(4) Restart the instrument after finishing the above inspections.

(5) If it still does not work correctly, please contact

RIGOL.

2. The signal is sampled but no waveform of the signal is displayed:

(1) Check whether the probe is correctly connected to the oscilloscope and

item to be tested.

(2) Check whether there are signals generated from the item to be tested (you

can connect the probe compensation signal to the problematic channel to

determine which has problem, the channel or the item to be tested).

(3) Resample the signal.

3. The tested voltage amplitude is greater or lower than the actual value

(this problem usually occurs when probe is used):

Check whether the probe ratio of the channel complies with the attenuation ratio

of the probe.

4. There is waveform display but not stable:

(1) Check the trigger source: press MENU (in the trigger control area

(TRIGGER) at the front panel)  Source to check whether the trigger

source selected complies with the signal channel actually used.

(2) Check the trigger type: general signals should use "Edge" trigger and video

signal should use "Video" trigger. Only when the proper trigger type is used,

can the waveform be displayed stably.

(3) Check the trigger level: adjust the trigger level to the middle of the signal.

(4) Change the trigger holdoff setting.

5. No display after pressing RUN/STOP:

Check whether the trigger mode in the trigger control area (TRIGGER) at the

front panel is "Normal" or "Single" and whether the trigger level exceeds the

waveform range. If yes, set the trigger level to the middle or set MODE to
"Auto".

Note:

Using AUTO could automatically finish the above setting.