Chapter 5 To Trigger the Oscilloscope

5-2

MSO1000Z/DS1000Z User’s Guide

Trigger Source

Press MENUSource in the trigger control area (TRIGGER) at the front panel to
select the desired trigger source. Analog channels CH1-CH4, digital channels D0-D15

or the AC Line can all be used as trigger source.

Analog channel input:

Signals input from analog channels CH1-CH4 can all be used as trigger source. No

matter whether the channel selected is enabled, the channel can work normally.

Note:

When any channel among D7-D0 is enabled, CH4 cannot be used as trigger

source; when any channel among D15-D8 is enabled, CH3 cannot be used as trigger

source.

Digital channel input:

Only the digital channels turned on can be used as trigger source. Please refer to the

introduction of "To Turn on/off the Digital Channel" to turn on the desired

digital channel.

AC line:

The trigger signal is obtained from the AC power input of the oscilloscope. AC trigger

is usually used to measure signals relevant to the AC power frequency. For example,

stably trigger the waveform output from the transformer of a transformer substation;

it is mainly used in related measurements of the power industry.