Chapter 4 To Set the Sample System

4-2

MSO1000Z/DS1000Z User’s Guide

Acquisition Mode

The acquisition mode is used to control how to generate waveform points from

sample points.

Press AcquireMode at the front panel and use

to select the desired

acquisition mode (the default is normal), then press down the knob to select this

mode. You can also press Mode continuously to switch the acquisition mode.

Normal

In this mode, the oscilloscope samples the signal at equal time interval to rebuild the

waveform. For most of the waveforms, the best display effect can be obtained using

this mode.

Peak Detect

In this mode, the oscilloscope acquires the maximum and minimum values of the

signal within the sample interval to get the envelope of the signal or the narrow pulse

of the signal that might be lost. In this mode, signal confusion can be prevented but

the noise displayed would be larger.

In this mode, the oscilloscope can display all the pulses with pulse widths at least as

wide as the sample period.

Average

In this mode, the oscilloscope averages the waveforms from multiple samples to

reduce the random noise of the input signal and improve the vertical resolution.

Greater number of averages can lower the noise and increase the vertical resolution;

while at the same time, it will slow the response of the displayed waveform to the

waveform changes.

When "Average" mode is selected, press Averages and use

to set the desired

number of averages. The number of averages can be set to 2, 4, 8, 16, 32, 64, 128,

256, 512 or 1024. The default is 2.