To Enable REF Function

Press REF in the vertical control area (VERTICAL) at the front panel to enable the
REF function.

MSO1000Z/DS1000Z provides 10 reference waveform channels. Press Channel and

use

to set the desired reference channel to on or off and the channel icon (for

example,

) of the channel enabled will be displayed at the left side of the screen

grid.

When the REF function is enabled, you can select different color for each reference

waveform, set the source of each reference channel, adjust the vertical scale and

position of the reference waveform and save the reference waveform to internal or

external memory as well as recall it when needed. For the details, please refer to the

introductions below.

To Select REF Source

Press Current and use

to select any of the reference channels (Ref1 to Ref10)

enabled and then press Source to specify a reference source (CH1-CH4, MATH or

D0-D15) for this channel.

To Adjust REF Waveform Display

You can adjust the reference waveform specified in Current.

Press REF to enable the REF function. Press Offset and use

to adjust the

vertical position of the reference waveform and press Scale and use

to adjust

the vertical scale of the reference waveform. At this point, you can also directly
rotate VERTICAL

POSITION and VERTICAL

SCALE to adjust the

vertical position and scale respectively.

To Save to Internal Memory

Press Save to save the waveform (screen region) in the specified source to the

internal memory as reference waveform.

Note:

This operation only saves the reference waveform in the volatile memory and

the waveform will be cleared at power-off.