Chapter 13 Signal Source

RIGOL

MSO1000Z/DS1000Z User’s Guide

13-9

Figure 13-9 Built-in Waveform-Haversine

3. Set the frequency

Press Frequency to set the frequency of the signal. For the setting method,

please refer to the introduction in "

Parameter Setting Method". The range is

from 100 mHz to 1 MHz.

4. Set the amplitude

Press Amplitude to set the amplitude of the signal. For the setting method,

please refer to the introduction in "Parameter Setting Method". When the

impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the

impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.

5. Set the DC offset voltage

Press Offset to set the DC offset of the signal. For the setting method, please

refer to the introduction in "

Parameter Setting Method". When the

impedance is HighZ, the range is from (-2.5 V + the current amplitude/2) to (2.5

V - the current amplitude/2); when the impedance is set to 50 Ω, the range is

from (-1.25 V + the current amplitude/2) to (1.25 V - the current amplitude/2).

6. Set the start phase

Press StartPhase to set the start phase of the signal. For the setting method,

please refer to the introduction in "Parameter Setting Method". The range is

from 0°to 360°.

7. Align Phase

Pressing AlignPhase will re-configure the two channels to output according to

the preset frequency and phase. For two signals of which the frequencies are the

same or are multiples, this operation can align their phases. Acquire the

waveforms of the two channels using the oscilloscope and display the waveforms

stably. After switching the channel status, the phase deviation between the two

waveforms is changed. At this point, press AlignPhase and the phase deviation

shown on the oscilloscope will restore to the current phase deviation between

the two waveforms automatically.