To create waveform, To create waveform -12, To create – RIGOL MSO/DS1000Z Series User Manual

Page 208: Waveform

Advertising
background image

RIGOL

Chapter 13 Signal Source

13-12

MSO1000Z/DS1000Z User’s Guide

Press SelectCHSignal to set the desired channel signal.

Press Source to select the waveform of any of the channels turned on

(CH1-CH4).

Press WaveRange to set the waveform in the "Cursor" or "Screen" region

for output. When "Cursor" is selected, Press CursorA or CursorB and use

to adjust the positions of the two orange cursor lines respectively; or

press CursorAB and rotate

to adjust the positions of cursors A and B

at the same time.

Press Probe to set the probe ratio when the analog channel signal is

recalled. You can select "0.1X", "0.2X", "0.5X", "1X", "2X" or "5X". For

example, if the selected analog channel amplitude is 10 VPP and the probe

ratio is "0.5X", the amplitude of the arbitrary waveform recalled is 5 VPP.

Pressing Load can recall the channel signal selected.

You can also edit the waveform currently selected; for the details, refer to the

introduction in "To Edit Waveform".

To Create Waveform

Users can create arbitrary waveforms according to their needs. Press Create to

enter the waveform creation menu.

1. Set the Number of Initial Points

The number of initial points refers to the number of editable points. By default,

the number of editable points of arbitrary waveform created can be preset to 2

points. Point 1 is fixed at 0 s and point 2 is fixed at the middle of the period. Press

InitPoint to set the number of editable points of the new waveform. For the

setting method, please refer to the introduction in "

Parameter Setting

Method". The arbitrary waveform can contain up to 16384 (16 kpts) points.

2. Interpolation

Press Interp to turn on or off the interpolation mode between the points of the

arbitrary waveform.


ON: the waveform editor connects two points using a straight line.