RIGOL MSO/DS1000Z Series User Manual

Page 90

Advertising
background image

RIGOL

Chapter 5 To Trigger the Oscilloscope

5-16

MSO1000Z/DS1000Z User’s Guide

"Don’t Care", the oscilloscope will not trigger.

or

: set the pattern to the rising or falling edge of the channel selected.

Note:

 When digital channels D7-D0 are enabled, CH4 is automatically disabled; the

corresponding pattern cannot be set and is replaced by X. When D15-D8 are

enabled, CH3 is automatically disabled; the corresponding pattern cannot be set

and is replaced by X.

 Only one edge (rising or falling edge) can be specified in the pattern. If one

edge item is currently defined and then another edge item is defined in another

channel in the pattern, the former edge item defined will be replaced by X.

All Bits:

Press AllBits to set the patterns of all trigger sources to the pattern settings selected

currently.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to "Trigger Mode". The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff and noise rejection)

under this trigger type.

Trigger Level:

For the analog channels, the trigger level of each channel needs to be set

independently. For example, set the trigger level of CH1. Press Source to select CH1,

and then use TRIGGER LEVEL to modify the level. For the details, please refer

to the introduction of "

Trigger Level".

Advertising