Chapter 2 To Set the Vertical System

RIGOL

MSO1000Z/DS1000Z User’s Guide

2-3

Bandwidth Limit

MSO1000Z/DS1000Z supports the bandwidth limit function which can reduce the

display noise. For example, the signal under test is a pulse with high frequency

oscillation.
 When the bandwidth limit is disabled, the high frequency components of the

signal under test can pass the channel.

 When limiting the bandwidth to 20 MHz, the high frequency components of the

signal under test that exceed 20 MHz are attenuated.

Press CH1, then press BW Limit continuously to switch the bandwidth limit state
(the default is OFF). When the bandwidth limit is enabled, the character "B" will be

displayed in the channel status label at the bottom of the screen.

Note:

Bandwidth limit can reduce the noise as well as attenuate or eliminate the

high frequency components of the signal.

Probe Ratio

MSO1000Z/DS1000Z allows user to set the probe attenuation ratio manually. Press

CH1Probe and use

to select the desired probe ratio. The probe ratio values

available are as shown in the table below.

Table 2-1 Probe Ratio

Menu

Attenuation Ratio

(display amplitude of the signal : actual

amplitude of the signal)

0.01X

0.02X

0.05X

0.1X

0.2X

0.5X

1X

2X

5X

10X (the default value)

20X

50X

100X

200X

500X

1000X

0.01:1

0.02:1

0.05:1

0.1:1

0.2:1

0.5:1

1:1

2:1

5:1

10:1

20:1

50:1

100:1

200:1

500:1

1000:1