Chapter 5 To Trigger the Oscilloscope

5-14

MSO1000Z/DS1000Z User’s Guide

 480P: the frame frequency is 60 frames per second; the TV scan line is 525;

progressive scan; the line frequency is 31.5 kHz.

 576P: the frame frequency is 60 frames per second; the TV scan line is 625;

progressive scan.

Trigger Mode:

Press Sweep to open the trigger mode list and select auto, normal or single. For the

details, please refer to "

Trigger Mode". The corresponding status light of the

current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameter (noise rejection) under this trigger type.

Trigger Level:

Use TRIGGER LEVEL to modify the level. For the details, please refer to the

introduction of "Trigger Level".

Tips
 For a better observation of the waveform details in the video signal, you can

set a larger memory depth first.

 In the trigger debugging process of video signals, the frequency in different

part of the signal can be reflected by different brightness as

RIGOL digital

oscilloscope provides the multi-level gray scale display function. Experienced

users can quickly judge the signal quality and discover abnormalities.