Slope trigger, Slope trigger -10 – RIGOL MSO/DS1000Z Series User Manual

Page 84

Advertising
background image

RIGOL

Chapter 5 To Trigger the Oscilloscope

5-10

MSO1000Z/DS1000Z User’s Guide

Slope Trigger

In slope trigger, the oscilloscope triggers on the positive or negative slope of the

specified time. This trigger mode is applicable to ramp and triangle waveforms.

Trigger Type:

Press Type, rotate

to select "Slope" and press down

. At this point, the

trigger setting information is displayed at the upper right corner of the screen.
For example,

. The trigger type is slope trigger; the trigger source

is CH1; the difference between the upper limit of trigger level and the lower limit of

trigger level is 400 mV.

Source Selection:

Press Source to switch the signal source and select CH1-CH4. For the details, please

refer to the introduction in "

Trigger Source". The current trigger source is

displayed at the upper right corner of the screen.

Note:

Select channel with signal input as trigger source to obtain stable trigger.

Slope Condition:

In this oscilloscope, positive slope time is defined as the time difference between the

two crossing points of trigger level line A and B with the rising edge; negative slope

time is defined as the time difference between the two crossing points of trigger level

line A and B with the falling edge as shown in the figure below.

Figure 5-4 Positive Slope Time/Negative Slope Time

Press When to select the desired slope condition.

: trigger when the positive slope time of the input signal is greater than

the specified time.

: trigger when the positive slope time of the input signal is lower than the

specified time.

: trigger when the positive slope time of the input signal is greater than

the specified lower limit of time and lower than the specified upper limit of time.

: trigger when the negative slope time of the input signal is greater than

the specified time.

: trigger when the negative slope time of the input signal is lower than

the specified time.

A

B

The upper limit of trigger level (Up Level)

The lower limit of trigger level (Low Level)

Positive

Slope Time

Negative

Slope Time

Advertising