To set the waveform display size, Reorder setting, Auto view – RIGOL MSO/DS1000Z Series User Manual

Page 160: To set the threshold, To set the waveform display size -4, Reorder setting -4, Auto view -4, To set the threshold -4

Advertising
background image

RIGOL

Chapter 7 Digital Channel

7-4

MSO1000Z/DS1000Z User’s Guide

To Set the Waveform Display Size

Press WaveSize to open the waveform size selection list. You can use

or press

WaveSize to set the waveform display size of the channel currently turned on to S
(small) and L (large).

Note:

L (large) can only be used when the number of channels currently turned on is

no more than 8.

You can also directly rotate VERTICAL

SCALE to set the waveform display

size. Turn clockwise to set it to L; turn counterclockwise to set it to S.

Reorder Setting

Press ReOrder to select the waveform order mode on the screen of the channels

currently selected. You can select "D0-D15" or "D15-D0" and the default is "D0-D15".


 D0-D15: the waveforms on the screen are D0-D15 from top to bottom.
 D15-D0: the waveforms on the screen are D15-D0 from top to bottom.

Auto View

Press Auto View and the analog channel waveform and digital channel waveform

will be displayed respectively on the upper and lower half screen respectively. By

default, the digital channel waveforms are displayed on the upper half screen and the

analog channel waveforms are displayed on the lower half screen.

Note:

When the number of channels currently turned on is more than 8 or the

waveform display size is L, Auto View is grayed out and disabled.

To Set the Threshold

Press Threshold to enter the threshold setting menu. When the voltage of the input

signal is greater than the threshold currently set, it is treated as logic 1; otherwise, it

is treated as logic 0.

The threshold levels of channels D7-D0 and channels D15-D8 can be set separately

according to the need. You can select a preset value or user-defined threshold. The

range of the user-defined threshold is from -15.0 V to +15.0 V.

Press Low Type to open the threshold selection list to select the threshold of D7-D0.

The selectable preset values include TTL (1.40 V), CMOS (2.50 V), ECL (-1.30 V),

PECL (3.70 V), LVDS (1.20 V) and 0V. You can also press D7-D0 and use

to set

Advertising