Chapter 10 Pass/Fail Test

RIGOL

MSO1000Z/DS1000Z User’s Guide

10-3

Test and Output

You can set the output form of the test results by the following method.

Press Stat.Disp to select "ON" or "OFF". When "ON" is selected, the test results will

be displayed at the upper right corner of the screen as shown in the figure below.

Press Stat.Reset to clear the current data and perform statistics on the test results

again.

Press StopOnFail to select "ON" or "OFF".
 ON: when failed waveforms are detected, the oscilloscope will stop the test and

enter the "STOP" state. At this point, if Stat.Disp is turned on, the last statistic
results will be displayed on the screen; if Aux Out is turned on, a pulse is output
from the [Trigger Out] connector at the rear panel.

 OFF: the oscilloscope will continue with the test even though failed waveforms

are detected. The test results on the screen will update continuously. If Aux
Out
is turned on, a pulse will be output from the [Trigger Out] connector at
the rear panel each time a failed waveform is detected.

Press Output to select "

" or " ".

: when failed waveforms are detected, the instrument displays the statistic

results and outputs a pulse.

: when failed waveforms are detected, the instrument displays the statistic

results and outputs a pulse and the beeper sounds (not related to the on/off

state of the sound).

Press AuxOut to quickly turn "ON" or "OFF" the output of test results from the

[Trigger Out] connector at the rear panel. When a failed waveform is detected, the

oscilloscope outputs a negative pulse from the [Trigger Out] connector at the rear

panel; when no failed waveform is detected, the instrument will output low level

continuously.