RIGOL MSO/DS1000Z Series User Manual

Page 37

Advertising
background image

Chapter 1 Quick Start

RIGOL

MSO1000Z/DS1000Z User’s Guide

1-11

Pass/Fail:

The instrument can output a negative pulse via this connector when a failed

waveform is detected during the pass/fail test. The instrument continuously

outputs a low level via this connector when no failed waveform is detected.

5 Source Output

The output terminals of the built-in dual-channel source of the oscilloscope.

When Source1 or Source2 is enabled, the signal currently set can be output

through the [Source1] or [Source2] connector at the rear panel.

6 Lock Hole

You can lock the instrument to a fixed location using the security lock (please

buy it yourself) via the lock hole.

7 Fuse

If a new fuse is required, please use the specified fuse (250V, T2A). The

replacing method is as follows.

a) Turn off the instrument, switch off the power supply and remove the power

cord.

b) Insert a small straight screwdriver into the slot at the power socket and pry

out the fuse seat gently.

c) Take out the fuse and replace it with a specified fuse. Then, install the fuse

seat.

8 AC Power Socket

AC power input terminal. The power requirements of this oscilloscope are

100-240 V, 45-440 Hz. Use the power cord provided with the accessories to

connect the instrument to AC power. Then, you can press the power key at the

front panel to start the instrument.