Chapter 1 Quick Start

RIGOL

MSO1000Z/DS1000Z User’s Guide

1-7

Function Inspection

1. Press StorageDefault to restore the oscilloscope to its default

configuration.

2. Connect the earth alligator clip of the probe to the "Ground Terminal" as shown

in the figure below.

3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the

"Compensation Signal Output Terminal" of the probe.

Figure 1-7 To Use the Compensation Signal

4. Press AUTO.
5. Observe the waveform on the display. In normal condition, the square waveform

as shown in the figure below should be displayed.

Figure 1-8 Square Waveform

6. Use the same method to test the other channels. If the square waveforms

actually shown do not match that in the figure above, please perform "Probe

Compensation".

WARNING

To avoid electric shock when using the probe, please make sure that the

insulated wire of the probe is in good condition and do not touch the

metallic part of the probe when the probe is connected to high voltage

source.

Tip

The signal output from the probe compensation connector can only be used for

probe compensation adjustment and can not be used for calibration.

Compensation Signal Output Terminal
Ground Terminal