Delay Calibration of the Analog Channel

When using an oscilloscope for actual measurement, the transmission delay of the

probe cable may bring relatively greater error (zero offset). MSO1000Z/DS1000Z

supports user to set a delay time to calibrate the zero offset of the corresponding

channel. Zero offset is defined as the offset of the crossing point of the waveform

and trigger level line relative to the trigger position, as shown in the figure below.

Figure 2-1 Zero Offset


Press CH1Delay-Cal and use

to set the desired delay time. The range

available is from -100 ns to 100 ns. Pressing down the multifunction knob

will

reset the delay time to 0.00 s.

Note:

This parameter is related to the instrument model and the current horizontal

time base setting. The larger the horizontal time base is, the larger the setting step

will be. Take MSO1104Z-S as an example. The step values under different horizontal

time base are as shown in the table below.

Table 2-2 Relation between delay calibration time step and horizontal time base

Horizontal Time Base Delay Calibration Time Step
5 ns

100 ps

10 ns

200 ps

20 ns

400 ps

50 ns

1 ns

100 ns

2 ns

200 ns

4 ns

500 ns

10 ns

1 μs to 10 μs

20 ns

Note:

When the horizontal time base is equal to or greater than 10 μs, the delay

calibration time cannot be adjusted.