RIGOL MSO/DS1000Z Series User Manual

Page 139

Advertising
background image

Chapter 6 MATH and Measurement

RIGOL

MSO1000Z/DS1000Z User’s Guide

6-25

Delay and Phase

Source 1

Source 2

Period

Delay

Figure 6-4 Delay and Phase

Source 1 and source 2, the same as source A and source B in the measurement

setting menu can be any channel of CH1-CH4 or D0-D15. For the setting method,

please refer to the introduction in "Measurement Setting".

1. Delay 12 : the time difference between the rising edges of source 1 and

source 2. Negative delay indicates that the selected rising edge of source 1

occurred after that of source 2.

2. Delay 12 : the time difference between the falling edges of source 1 and

source 2. Negative delay indicates that the selected falling edge of source 1

occurred after that of source 2.

3. Phase 12 : phase difference calculated according to "Delay 1→2 " and the

period of source 1, expressed in degree. The calculation formula is as shown in

equation (6-1).

4. Phase 12 : phase difference calculated according to "Delay 1→2 " and the

period of source 1, expressed in degree. The calculation formula is as shown in

equation (6-1).

The phase calculation formula:

× 360

Delay

Phase =

Period1

°

(6-1)

Wherein,

Phase

represents "Phase 12 " or "Phase 12 "

Delay

represents "Delay 12 " or "Delay 12 "

1

Period

represents the period of source 1

Advertising