<3. INSTALLATION>

3-2

IM 01E30D01-01EN

(5) Precautions for Use of Liquid Sealing

Compounds

IMPORTANT

Care must be taken in using liquid sealing

compounds on the piping, as it may have a

negative influence on the flow indications by

flowing out and covering the surfaces of an

electrode or grounding ring. In particular, care

must be taken if a liquid sealing compound is

used in the case of vertical piping.

(6) Service Area
Select locations where there is adequate space to

service installing, wiring, overhauling, etc.

(7) Bypass Line
It is recommended to install a bypass line to facilitate

maintenance and zero adjustment.

Bypass valve

Block valve

Block valve

F0303.EPS

Figure 3.1.3 Bypass Line

(8) Supporting the Flowmeter

CAUTION

Do not secure the flowmeter separately to

prevent the vibrations, shocks, and expansion

and contraction forces of the piping from

affecting it. Fix the pipes first, then support the

flowmeter with the pipes.

(9) Mounting Positions

• Pipes must be fully filled with liquids.

IMPORTANT

If the pipe is empty, the output fluctuates or the

Process Alarm (Signal Overflow) occurs. The

pipe must be fully filled with liquid.

Piping shall be designed so as to maintain the

interior of the flowtube filled with fluids.

Vertical mounting is effective in such cases as

when fluids tend to separate or solid matter may be

precipitated. When employing vertical mounting,

direct the fluids from the bottom to the top to ensure

that the pipes remain fully filled.

F0304.EPS

(Incorrect)

(Correct)

(Incorrect)

(Correct)

h

h>0

h>0

h

Figure 3.1.4 Mounting Positions

• Avoid air bubbles.

IMPORTANT

If air bubbles enter a measurement pipe,

flow rate indications may be affected and

measurement errors may be caused.

In cases where fluids contain air bubbles, piping must

be designed to prevent them from accumulating in

the measurement pipe of a flowtube.

If a valve exists near the flowmeter, try to mount the

flowmeter on the valve’s upstream side in order to

prevent a possible reduction of pressure inside the

pipe, thereby avoiding the possibility of air bubbles.

Valve

(Incorrect)

(Correct)

(Incorrect)

(Correct)

F0305.ai

Figure 3.1.5 Avoiding Air Bubbles