<6. PARAMETER DESCRIPTION>

6-39

IM 01E30D01-01EN

[N15: Test Time] Selection of the holding time of

test mode

This parameter is used to select the holding time of

test mode from the last operation in test mode.

Setting

Function

10min

The holding time is 10min

30min

The holding time is 30min

60min

The holding time is 60min

3h

The holding time is 3h

6h

The holding time is 6h

12h

The holding time is 12h

[N20: Test Iex] Selection of the test for excitation

current

This parameter is used to select a test for excitation

current. The excitation current value is forcibly fixed

at zero during this test. Setting is possible when

“Test” has been selected for

N10: Test Mode.

Setting

Function

Normal

Normal operation

Test

(Iex OFF)

Testing is started.

The excitation current value is forcibly fixed at

zero.

IMPORTANT

1) The default setting is “Normal.”

Use the instrument with this default setting.

2) When the setting is “Test (lex OFF),” the

output is fixed to 0%.

In the case of setting “Test (lex OFF),” first

change the control loop to the manual mode,

and then set this parameter.

3) “Normal” will be restored when the power is

turned off or when the time set at

N15: Test

Time have elapsed since entry to “Test (lex

OFF).”

[N30: Flow Tube] Selection of operating flowtube

type

This parameter is used to select the type of

operating flowtube which is connected to the

converter. The parameter would be selected

as “Detector” for typical use. This parameter is

selected as “Calibrator” when the converter is

connected to the calibrator.

[N40: Average Execution] Selection of execution

for measurement of averaged flowrate

This parameter is used to select whether the

execution for measurement of averaged flowrate is

occurrence in a constant time.

Setting

Function

Stop

Stopping the execution of measurement for

averaged flowrate.

Start

Starting the execution of measurement for

averaged flowrate.The totalization is started in

a time, which was set at

N41: Average Timer

using the value being set at

B20(C11): Flow

Damping when this function is started. This

totalization is calculated at the result of instant

flowrate for a time where was set at

N41.

[N41: Average Timer] Setting of totalization time

for measurement of averaged flowrate

This parameter is used to set the totalization time

for averaged flowrate measurement.

[N42: Average Flow] Display of the averaged

flowrate

This parameter is used to display the averaged

flowrate for a time when is set at

N41 as totalizatoin

time after selecting the “Start” at

N40.

(12) Menu P: Parameter Protection items
Menu P contains items that are relevant to write

protection and passwords.

[P10: Key Code] Parameter of the display

restriction

This parameter restricts access to the Service

Mode.

NOTE

Write Protect function

■ The parameters

P20 through P23 are

set when using the write protect function.

Specifically, this function responds to a

hardware switch or the setting of a software

password, and it protects parameters from

being overwritten.

■ If the hardware switch is set to “Protect”, it

will not be possible to overwrite parameters;

furthermore, this condition will be maintained

until the switch is set to “Enable.”

■ For more details regarding hardware switch

settings, refer to Section 11.3.2: Setting of

Write Protect Switch.