<5. BASIC OPERATING PROCEDURES

(USING THE DISPLAY UNIT)

>

5-1

IM 01E30D01-01EN

5. BASIC OPERATING PROCEDURES

(USING THE DISPLAY UNIT)

Both of four push button and magnet switches can

be used for setting. The magnet switches can be

used to set parameters without opening the case

cover.
The magnet switches need operational magnets

(optional code BM). These are also available as

part number F9840PA.
Users can choose a language on indicator from

English, Japanese, German, French, Italian, and

Spanish.
This chapter will provide a description of basic

data configuration and the methods to be used

with the four setting switches. The AXR can also

be operated using a configuration tool (Such

as handheld Brain Terminal (BT200), a HART

Configuration Tool or FieldMate, etc.). (Please

refer to Chapter 7 for operation via Brain Terminal

and Chapter 8 and 9 for operation via HART

Configuration Tool.

WARNING

Be sure to set parameters as “Protect” on the

write protect function after finish of parameter

setting work.

Refer to “Chapter 6 Menu P: Parameter

Protection items” and section “11.3.2” how to use

the write protect function in detail.

CAUTION

When using this instrument for a long time in

conditions of high temperature and humidity, the

visibility of the display unit may deteriorate. In

this case, it is necessary to replace the display

unit.

CAUTION

Precautions for Push Button Switches

• Open the cover when operating push button

switches. Refer to Section 10.2 for details on

opening the cover.

• Do not open the cover in wet weather or high

humidity in order to prevent condensation.

The push button switches cannot be used

in these conditions as the cover would need

to be opened. In this case, use magnet

switches to set parameters.

• Do not touch the push button switches with

wet or dirty hands.

NOTE

Precaution for Magnet Switches and

Operational Bar-Magnet

(1) A bar-magnet with its case is attached

with an AXR when the optional code BM is

chosen at the same time. This bar-magnet

with its case is also available as part number

F9840PA.

(2) The bar-magnet should be used for the

parameter setting of AXR from its outside

glass window only.

(3) The switch operation should be done under

the condition of closing the cover case

outside its glass window.

(4) Do not hit the bar-magnet to the glass of the

cover so strongly.

(5) If dirt, dust or other substances surfaces on

the display panel glass, wipe them clean with

a soft dry cloth.

(6) Do not make a condition of closing the

bar-magnet to the flowtube of AXR in order

to prevent giving an influence of magnetic

distribution inside the flowtube and of the

accuracy of AXR.

(7) Do not bring the bar-magnet near equipment

susceptible to magnetic field such as

magnetic cards, medical equipment, etc.

(8) Keep the bar-magnet in the case because of

the powerful magnetic force.