<6. PARAMETER DESCRIPTION>

6-13

IM 01E30D01-01EN

(10) Item K (Menu K): Diagnostic Function Setting items
Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.

Item

Name

R/W

Data range

Units

Position

of decimal

point

Default value

(*): Indicated item

Description

Display unit

(BRAIN)

Display unit

/BRAIN

K00 Diagnosis

(DIAGNOSIS)

K10 Adhesion Check

(ADHESION CHK)

W

No Execution

Execution

No Execution

Selects whether or not to

perform diagnosis of adhesion

to the electrode.

K11 Adh Status

(ADH STATUS)

R

Level0

Level1

Level2

Level3

Level4

Level0

Displays the status of adhesion

level

K12 Adh Measure Value

(ADH MEAS VAL)

R

M ohm

2

0.00M ohm

Displays the resistance value

for adhesion to the electrode.

K13 Adhesion Level1

(ADH LEVEL1)

W

0.00 to 100.00

M ohm

2

0.35

Sets the resistance value

for adhesion Level 1 to the

electorode.

K14 Adhesion Level2

(ADH LEVEL2)

W

0.00 to 100.00

M ohm

2

0.40

Sets the resistance value

for adhesion Level 2 to the

electorode.

K15 Adhesion Level3

(ADH LEVEL3)

W

0.00 to 100.00

M ohm

2

0.45

Sets the resistance value

for adhesion Level 3 to the

electorode.

K16 Adhesion Level4

(ADH LEVEL4)

W

0.00 to 100.00

M ohm

2

0.60

Sets the resistance value

for adhesion Level 4 to the

electorode.

K17 Adh Chk Cyc

(ADH CHK CYC)

W

0.5 min

1 min

5 min

10 min

min

5 min

Selects execution time of

adhesion check

K19 Empty Check

(EMPTY CHECK)

W

No

Low

Middle

High

No

Selects whether or not to

perform diagnosis of empty

pipe check.

K20 Empty Status

(EMPTY STATUS)

R

Normal

Empty

Normal

Displays status of empty pipe

K21 DC Voltage A

(DC VOLTAGE A)

R

V

2

Displays the voltage level at

electrode A

K22 DC Voltage B

(DC VOLTAGE B)

R

V

2

Displays the voltage level at

electrode B

K23 Empty Level

(EMPTY LEVEL)

W

-1.5 to 1.5

V

1

-1.2V

Sets the volatage level for

empty pipe detection at the

electrodes

K60 —

(SELF CHECK)

R

Good

Error

See “6.5 Alarm Functions”.