<3. INSTALLATION>

3-4

IM 01E30D01-01EN

(4) Terminal Box Cover

IMPORTANT

As it is possible that the insulation will

deteriorate, do not open the terminal box cover

until it is time to wire it.

(5) Long-term Non-use

IMPORTANT

It is not desirable to leave the flowmeter unused

for a long term after installation. If this situation

is unavoidable, take care of the flowmeter by

observing the following.

• Confirmation of sealing conditions for the

flowmeter

Confirm that the terminal box screw and wiring ports

are well sealed. Equip the conduit piping with drain

plugs or waterproof glands to prevent moisture or

water from penetrating into the flowmeter through

the conduit.

• Regular inspections

Inspect the sealing conditions as mentioned above,

and the inside of the terminal box at least once a

year. Also, due to rain, etc. when it is suspected

that water may have penetrated into the inside

flowmeter perform supplementary inspections.

3.2.2 Flowmeter Piping

CAUTION

Misaligned or slanted piping can lead to leakage

and damage to the flanges.

(1) Correct any misaligned or slanted piping, and

any gaps that may exist between mounting

flanges before installing the flowmeter (refer to

Figure 3.2.1).

Slanted

Misaligned

F0307.ai

Figure 3.2.1 Slanted and Misaligned Flowmeter

Piping

(2) Inside a newly installed pipeline, there may

be some foreign substances such as residue

from welding or wood chips. Remove them

by flushing the piping before mounting the

flowmeter. This prevents the lining from

being damaged, as well as the occurrence

of erroneous measured signals resulting

from foreign substances passing through the

flowtube during measurement.