Maintenance, 1 changing direction ofelectrical connection, 1 changing direction of electricalconnection – Yokogawa ADMAG AXR User Manual

Page 141: Maintenance -1, Changing direction of electrical connection -1, Warning, Important, Caution, 1 changing direction of electrical connection

Advertising
background image

<11. MAINTENANCE>

11-1

IM 01E30D01-01EN

WARNING

• Maintenance work must be carried out by

the trained personnel having knowledge of

safety standard and not by operators.

• Do not open the cover in wet weather or high

humidity.

• When opening the cover, wait for more

than 5 minutes after turning off the power.

Furthermore, opening of the cover must

also be carried out by the trained personnel

having knowledge of safety standard.

IMPORTANT

• The amplifier assembly contains sensitive

parts that may be damaged by static

electricity. Exercise care so as not to directly

touch the electronic parts or circuit patterns

on the board, for example, by preventing

static electrification by using grounded wrist

straps when handing the assembly. Also

take precautions such as placing a removed

amplifier assembly into a bag with an

antistatic coating.

CAUTION

• Explosion protected type must be, as a

rule, removed to a non-hazardous area for

maintenance and be disassembled and

reassembled to the original state.

• The cover is locked by the special screw.

In case of opening the cover, use the

hexagonal wrench (nominal size 3).

• Be sure to lock the cover by the special

screw using this hexagonal wrench after

installing the cover.

11.1 Changing Direction of

Electrical Connection

11.1.1 Changing Direction of Electrical

Connection

(1) The following tools are required to change the

direction of the electrical connection:

• Hexagonal wrench (nominal size 2.5)

• Wrench

(2) Turn off the power to the flowmeter.

(3) Using the wrench, loosen the hexagonal nut at

the neck of the instrument.

F1101.ai

(4) Using the hexagonal wrench (nominal size 2.5),

loosen the screw in the neck.

(5) Turn the converter in the desired direction.

NOTE

The converter can be turned –140 degree to

+180 degree from the arrow mark indicating the

flow direction. Do not exceed these angle.

(6) Using the hexagonal wrench (nominal size 2.5),

retighten the neck screw.

F1102.ai

(7) Using the wrench, retighten the hexagonal nut

at the neck. After that, check that the converter.

11. MAINTENANCE

Advertising