<6. PARAMETER DESCRIPTION>

6-46

IM 01E30D01-01EN

Setting Alarms (Device and measurements are normal but a warning is issued.)

Display unit/

BRAIN(60) content

Alarm countermeasure

message on display unit

NE-107

status

Alarm description

Countermeasure

80:Adhesion Wng

M

Slight adhesion to electrodes. Clean and check the electrodes.

Refer to parameter K15.

82:Auto Zero Wng

C

Results of automatic zero

adjustment are higher than

the rated values.

Carry out adjustment as follows:

· Check if the flowtube is filled with fluid.

· Check if the flow velocity is completely

zero.

· Check the condition of grounding.

83:Fix Cur Wng

S

The current value is fixed.

Confirm whether the flow rate is in

excess of the upper limit (a set value

at

J11) or below the lower limit (a set

value at

J12), or whether upon entry to

the Test Mode or not.

84:Disp Over Wng

(only for display unit)

C

Overflow in the display digits

during instantaneous flow

rate display.

Check whether parameter C43 is

correct.

90:Disp SW Wng

(only for display unit)

Display unit switches are not

operating.

Replace the display unit.

91:Disp Cur Wng

F

Excessive current is used on

the display unit.

Replace the display unit.

93:Dev Sim Wng

C

In HART simulation

Check whether the device is in HART

Simulation.

6.5.4 NE-107 status

When

[G35: Alarm Display] is set in “NE107”, AXR alarms are classified into 4 status categories. The NE-107

status is shown in the display unit and configuration tools.

NE-107 status

Status of the device

F

Failure

Device malfunction, Parts malfunction

C

Function Check

The output signal is temporarily invaild for the local operation or manual operation.

S

Out of specification

The device works in out of specification.

The output signal is uncertain for the process or the ambience.

M

Maintenance required

The maintenance is required in the near future.

Upon occurrence of an alarm

NE-107 Status

Upon issuance of a warning

NE-107 Status

F0621.ai

Process Alarm

S30:Sig Overflow

M

Check signal cable

and grounding