<6. PARAMETER DESCRIPTION>

6-40

IM 01E30D01-01EN

[P20: Write Protect] Setting of password to release

the write protection function

This parameter is used to indicate whether or not

write protection is currently on.

F0615.ai

Default setting (Enable)

P20:Write Protect

No

P20:Write Protect

Yes

F0616.ai

Write protection (Protect)

NOTE

The mark of “ ” is displayed at the right

periphery of the first line in the “Major item

parameter search mode” and “Sub-item

parameter search mode” during acting the write

protection.

[P21: Enable Wrt Passwd] Setting of password to

release the write protection function

When the correct password is input, write

protection will be released for a period of 10

minutes; furthermore, this period will be extended

by a further 10 minutes each time a parameter is

overwritten.

P21:Enable Wrt Passwd

F0617.ai

The cursor will flash when entering Parameter

Replacement Mode, and the password set with

P22: New Password should be input at this time.

[P22: New Password] Setting of a new password

This parameter sets the password required for

the release of write protection. When set, it will

be possible to make write protect settings on the

software side.

P22:New Password

F0618.ai

Default setting

The default setting for this parameter is a string of

8 spaces (i.e., Enable), and thus, the password

field will be empty. When the cursor is flashing, the

password should be input. Press the SET key twice

to confirm the password. The display will go back to

“Sub-item parameter search mode.”

F0619.ai

After password setting

P:Protect

21:Enable Wrt Passwd

22:New Password

23:Software Seal

To change a password, first of all use the password

originally set with

P21: Enable Wrt Passwd to

release the write protect function, and then set the

new password. Alternatively, if it is desired to return

to the condition where no password is set, enter a

string of 8 spaces.

[P23: Software Seal] Display the software seal

When the joker password has been used to

release write protection, this parameter displays

“Break,” and when protection is cancelled using the

password set using

P22: New Password, it returns

to “Keep.”

NOTE

If you should forget your password, the joker

password can be used to temporarily release

write protection function. To obtain the joker

password, please contact your nearest

YOKOGAWA sales office.