<6. PARAMETER DESCRIPTION>

6-22

IM 01E30D01-01EN

(3) Menu D: Total Setting items
Menu D contains parameters that are relevant to

totalization function settings.

[D10: Total Unit] Setting of units for totalization

scale

Refer to the description of parameter

B30.

[D11: Total Scale] Setting of the totalization scale

Refer to the description of parameter

B31.

[D12: Total Decimal Pnt] Setting of the decimal

point position for the totalization display

This parameter sets the position of the decimal

point for totalization display in terms of the number

of digits. Except in cases where 0 is selected, the

totalized units are not displayed.

Example: When totalized value is 12345678 m

3

Item

Totalization display

0

12345678 m

3

1

1234567.8

2

123456.78

3

12345.678

4

1234.5678

5

123.45678

6

12.345678

7

1.2345678

[D13: Total Low Cut] Setting of the totalization stop

range

This parameter allows the settings to be made that

prevent totalization when the flow rate is at or below

the low-cut setting value. The low cut setting value

depends on flow rate span.

● Flow rate span < 1 m/s:

Input value or 3 cm/s, whichever is bigger

● Flow rate span ≥ 1 m/s:

Input value

To confirm the flow rate span velocity in ”m/s” unit,

see

[C44: Velocity Check].

If the multiple ranges or forward/reverse ranges is

used, low cut is carried out at the setting value for

the smallest span.

Example: Situation where “the first range” is set to

“0.8 m/s” and “the second range” is set to

“1.5 m/s”

- When “5 %” is input in this parameter:

No totalization is carried out at flow

rates of 4 cm/s or lower.

- When “3 % (i.e. 2.4cm/s)” is input in this

parameter:

No totalization is carried out at flow

rates of 3 cm/s or lower.

However, the low cut function will be terminated if

this parameter is set to “0 %”.

NOTE

When the low cut is small, an incorrect output

may occur at the flow rate of zero and the

totalization may be carried out. Use the

instrument with the default setting 3%. If the span

is small, the damping time is short, or the fluid is

low conductivity, the totalization may be carried

out easily at the flow rate of zero. In this case, set

the span to be large, set the damping time to be

long, or set the low cut to be large.

[D14: Total Rate Check] Display of the totalization

rate in “Pulse/s”

This parameter displays the totalization rate, which

is converted to “Pulse/s” while 100% of the range

flows.

When the multiple ranges or forward/reverse

ranges are used, the AXR displays the totalization

rate which 100% of their maximum range flows.

[D20: Total Execution] Operation setting for the

totalization function

This parameter sets “Start” and “Stop” of the

totalization function, in addition to performing the

preset function for the forward totalized value and

the reverse totalized value.

*: The preset function starts the count for totalization from the set

value.

Item

Description

Start

(initial value)

Starts totalization

Stop

Stops totalization

Preset Total Sets the preset value for totalization display

that has been specified as the forward totalized

value. Preset value are determined using

D21:

Ttl Set Val Lower and D22: Ttl Set Val Upper.

In case that “Start” is sellected, the count for

totalization starts from the preset value. Setting

of zero as the preset value allows the zero-reset

function to be implemented.

Preset Rev

Total

Sets the preset value for totalization display

that has been specified as the reverse totalized

value. Preset value are determined using

D21:

Ttl Set Val Lower and D22: Ttl Set Val Upper.

In case that “Start” is sellected, the count for

totalization starts from the preset value. Setting

of zero as the preset value allows the zero-reset

function to be implemented.

[D21: Ttl Set Val Lower] Setting of the totalization

preset value (lower 6 digits)

This parameter sets a totalization preset value

in the lower 6 digits of the 8-digit totalized value.

If zero is to be set as the preset value, “000000”

should be set here.