Yokogawa ADMAG AXR User Manual

Page 46

Advertising
background image

<5. BASIC OPERATING PROCEDURES

(USING THE DISPLAY UNIT)

>

5-7

IM 01E30D01-01EN

5.3.2 Setting Example for Numeric-Type Data: Flow rate span

This example describes the setting of the flow rate span for the numeric-type parameter

B23: Flow Span from

100 l/min to 120 l/min.

Start:

Major Item

Parameter

Search Mode

F0505-1.EPS

Setting mode

X4

Major item

parameter

Major item

Sub-item

parameter

Major item

Sub-item

selection (B)

X4

X2

Parameter

Replacement Mode

Sub-item

Parameter

Search Mode

Setting mode

P:Protect

B:Easy Setup

C:Basic Setup

B:Easy Setup

50:Auto Zero Exe

10:Language

20:Flow Damping

B:Easy Setup

22:Base Time Unit

23:Flow Span

24:Flow Decimal Pnt

Selection of the appropriate parameter

B23:Flow Span

100 l/min

0000100. l/min

Rng:0.00001 32000

INC

SET

B23:Flow Span

100 l/min

000100. l/min

Rng:0.00001

32000

SET

INC

DEC /SHIFT

ESC

SET

DEC

/SHIFT

INC

SET

ESC

INC

DEC/SHIFT

Setting Mode Condition
Touch the

SET

switch to access

B: Easy Setup.

NOTE

The ▲ and ▼ symbols to the left of the

parameters indicate that additional setting items

to those being currently displayed may also be

selected. Use the

INC

switch to cycle through

these items.

Sub-Item Parameter Search Mode has been

accessed in this screen.

Touch the

INC

switch to move the cursor to

B23:

Flow Span.

The cursor has been moved to

B23: Flow Span in

this screen. (Sub-item selection screen (B))

Touch the

SET

switch to access Parameter

Replacement Mode.

Once Parameter Replacement Mode has been

selected, the digit that can be replaced will be

flashed on and off. When in this condition, confirm

the relevant setting range as displayed at the

bottom of the screen and then set the parameter as

required.

In this example, the parameter will be set to “120 l/min.”

NOTE

When setting a new value, use the

DEC /SHIFT

switch to move from digit to digit, and use the

INC

switch to cycle through values for each

individual digit. In addition to digit, it is also

possible to select a decimal point, and this allows

the position of the decimal point to be changed.

Modify the value to “120 l/min” as follows:

Touch the

DEC /SHIFT

switch to move the cursor

to the position for multiples of 10. Then, touch the

INC

switch to change the value at this position

from “0” to “2.”

When the value of “120” has been setup, touch the

SET

switch.

Advertising