1) display and output condition for system alarms, 2) display and output condition for process alarms – Yokogawa ADMAG AXR User Manual

Page 90

Advertising
background image

<6. PARAMETER DESCRIPTION>

6-42

IM 01E30D01-01EN

(1) Display and output condition for system alarms

Alarm description

Alarm

Output

4-20mA

output

Totalization

Pulse

Display unit

Alarm

record

Normal

Closed

(ON)

Normal

Normal

Normal

Display Mode No

10 μP Fault

Microprocessor

(CPU) failure

Open

(Off)

Fixed at

3.2mA or less

otherwise

21.6mA or

more (*)

Indetermination Indetermination Indetermination Indetermination

11 EEPROM

Fault

EEPROM failure

Open

(Off)

Fixed at

3.2mA or less

otherwise

21.6mA or

more (*)

Stopped

Stopped

Alarm Mode

(display of

system Alarm

message)

Recorded

12 Sub uP Fault Substitute

mcroprocessor

(CPU) failure

13 EX Pwr Fault Extitation power

circuit failure

14 A/D(S) Fault

A/D converter failure

(Signal)

15 A/D(I) Fault

A/D converter failure

(Excitation current)

16 Analog Fault Analog circuit failure
17 Coil Open

Flowtube coil is

open-circuit

18 Coil Short

Flowtube coil is

short-circuit

19 Excite Error

Excitation circuit

failure

20 Pulse Error

Discrepancy of

pulse count

21 EEPROM

Dflt

EEPROM default

values

Note: • Alarm is output only when

[F10:DO Function] is set to “Alarm Output.”

• The operation when “Open (Off) Act” is set for

[F11:DO Active Mode] is shown in above table.

*: The output value is performed in accordance with the setting of the burnout switch. For information about this switch, see Section 10.3.1.

(2) Display and output condition for process alarms

Alarm description

Selection

(parameter

number)

Alarm

output

4-20 mA

output

Totalization

Pulse

output

Displayunit Alarm

record

30 Sig Overflow Input signal error

YES

(G26)

Open

(Off)

Fixed

Stopped

Stopped Alarm Mode

(Message)

Recorded

NO

(G26)

Closed

(On)

Continuous

(*)

Continuous

(*)

Continuous

(*)

Display

Mode

No

31 Empty Pipe

Flowtube is not filled

with fluid

YES

(G27)

Open

(Off)

Fixed

Stopped

Stopped Alarm Mode

(Message)

Recorded

NO

(G27)

Closed

(On)

Continuous

(*)

Continuous

(*)

Continuous

(*)

Display

Mode

No

32 H/L Alm

H/L Alarm

YES

(G28)

Open

(Off)

Normal

operation

Normal

operation

Normal

operation

Alarm Mode

(Message)

No

NO

(G28)

Closed

(On)

Display

Mode

33 Adhesion Alm Electrode adhesion

alarm

YES

(G29)

Open

(Off)

Fixed

Stopped

Stopped Alarm Mode

(Message)

Recorded

NO

(G29)

Closed

(On)

Continuous

(*)

Continuous

(*)

Continuous

(*)

Display

Mode

No

Note: • larm is output only when

F10: DO Function is set to “Alarm Output.”

• The operation when “Open (Off) Act” is set for

F11: DO Active Mode is shown in above table.

• 4-20mA output upon the occurrence of an alarm will be fixed at the value selected with

G25: 4-20mA Process Alarm.

*: Although outputs are continuous, output values are not guaranteed.