<6. PARAMETER DESCRIPTION>

6-20

IM 01E30D01-01EN

(2) Menu C: Basic Setting items
Menu C principally contains the basic setting items

for the flowtube.

NOTE

In order to ensure that correct flow rate data can

be acquired, it is crucial that the nominal size,

flow rate span, and meter factor of the combined

remote flowtube are set. In cases where the

AXR integral flowmeter is ordered, the nominal

size and meter factor will be set upon shipment

from the manufacturing plant, and these will not

require additional setting.

If a flow rate span was specified upon ordering,

this will be set before shipment. If this is not

the case, however, it will be necessary for the

appropriate value to be set by the user.

[C10: Tag No] Setting of the tag number

The setting for this parameter corresponds to

one of the ordered items.

Up to a maximum of 16 characters can be entered

for the display unit. In case of using the HART

configuration Tool, up to a maximum of 8 characters

can be entered. For more details regarding the

actual characters that can be used, refer to Section

5.2.2: Setting Mode.

NOTE

If a tag number is specified upon ordering, this

parameter is set up before shipment; however,

if this is not the case, it will be necessary for the

appropriate value to be set up by the user.

[C11: Flow Damping] Setting of the damping time

Refer to the description of parameter

B20.

[C20: Low MF (IEL)] Setting of the low-frequency

meter factor (Excitation current (Low))

This parameter sets the low-frequency meter factor

for excitation current (Low).

[C21: High MF (IEL)] Setting of the high-frequency

meter factor (Excitation current (Low))

This parameter sets the high-frequency meter

factor for excitation current (Low).

[C22: Low MF (IEM)] Setting of the low-frequency

meter factor (Excitation current (Middle))

This parameter sets the low-frequency meter factor

as required for excitation current (Middle).

[C23: High MF (IEM)] Setting of the high-frequency

meter factor (Excitation current (Middle))

This parameter sets the high-frequency meter

factor for excitation current (Middle).

[C24: Low MF(IEH)] Setting of the low-frequency

meter factor (Excitation current (High))

This parameter sets the low-frequency meter factor

for excitation current (High).

[C25: High MF(IEH)] Setting of the high-frequency

meter factor (Excitation current High))

This parameter sets the high-frequency meter

factor for excitation current (High).

NOTE

Meter Factor Settings

Meter factors

F0601.ai

42

(1) The values of the meter factors which are

marked at METER FACTOR fields on the

data plate of the AXR are set in the AXR at

the manufacturing plant.

(2) The meter factors are crucial in ensuring

that the electromotive force is correctly

in proportion to the flow velocity and are

determined at the manufacturing plant by

actual-flow calibration.

[C30: Nominal Size Unit] Setting of the nominal

size units

This parameter selects the units used for setting of

the nominal size.

[C31: Nominal Size] Setting of the nominal size

This parameter sets the nominal size of flowtube.

[C40: Base Flow Unit] Selection of flow units for

the flow rate span

Refer to the description of parameter

B21.