Yokogawa ADMAG AXR User Manual

Page 39

Advertising
background image

<4. WIRING>

4-15

IM 01E30D01-01EN

Table 4.6.7 TIIS Explosion Type Wiring Examples (Digital External Indicator Using Current Output)

Connection

Description

Current Output

Example 1

Connection to digital

external indicator

Distributor, etc.

+

R3

E1

Cable Resistance:R1[Ω]

*9

Cable Resistance:R4[Ω]

*9

0.0236 × (R1 + R2 + R3 + R4) + V2 + 14.7 ≤ E1 [V] ≤ 42

*10

In the case of R3 = 250 [Ω]

0.0236 × (R1 + R2 + R4) + V2 + 20.6 ≤ E1 [V] ≤ 42

*10

Cable Resistance:R2[Ω]

*9

Digital External Indicator

Relaying

terminal

77.7%

V2: Voltage between terminals of External Indicator

V2

R3

[

Ω

]:Load Resistance

JIS Class A

Hazardous area Non-hazardous area

AXR Terminal

SUPPLY

DO

Transmission of AXR

output might fail when

the external indicator

breakdown occurs.

+

-

+

-

+

-

-

Example 2

Connection to analog

external indicator

Cable Resistance:R4[Ω]

*9

Distributor, etc.

R3

E1

0.0236 × (R1 + R2 + R3 + R4 + 10) + 14.7 ≤ E1 [V] ≤ 42

*10

In the case of R3 = 250 [Ω]

0.0236 × (R1 + R2 + R4 + 10 ) + 20.6 ≤ E1 [V] ≤ 42

*10

Analog External Indicator

Internal resistance 10 Ω or less

R3

[

Ω

]:Load Resistance

JIS Class A

AXR Terminal

SUPPLY

DO

Relaying

terminal

Transmission of AXR

output might fail when

the external indicator

breakdown occurs.

+
-

+
-
+
-

+

-

Cable Resistance

:

R1[Ω]

*9

Cable Resistance

:

R2[Ω]

*9

Hazardous area Non-hazardous area

*6:

To avoid the influence of external noise, use an electric counter which fits to the pulse frequency.

*7:

Resistor is not necessary in the case of an electric counter which can receive contact pulse signal directly.

*8:

Ground the AXR to avoid the current output error in simultaneous current-pulse output.

*9:

Calculate the cable resistance by using the following as a rough guideline: 10.9 Ω per 1 km for the cable with the cross section of

2 mm

2

, 19.5 Ω per 1 km for the cable with the cross section of 1.25 mm

2

.

*10: The maximum voltage is 32 V DC in the case of Lighting Protector specification (optional code A).

(5) Installing the Cover
Install the cover to the flowmeter by turning it in the direction of the arrow as shown below. Tighten cover

locking screws counterclockwise using a hexagonal wrench (nominal size 3) to lock the cover (in case of

explosion proof type).

F0418.ai

Explosion proof type

Both Explosion proof and

General purpose types

Cover locking screw

Front

side

Figure 4.6.5 Installing the Terminal Box Cover