<6. PARAMETER DESCRIPTION>

6-2

IM 01E30D01-01EN

6.3 Parameter List Overview

(1) Item A (Menu A): Display items
Menu A contains the instantaneous flow rate, totalization values, and other items relevant to display.

Item

Name

R/W

Data range

Units

Position

of decimal

point

Default value

(*): Indicated item

Description

Display unit

(BRAIN)

Display unit

/BRAIN

A00 Display

(DISPLAY)

A10 FR

(FLOW RATE (%))

R

-103.1 to 103.1

%

1

For Display Mode only

A20 FR

(FLOW RATE)

R

-999999 to 999999 B21/B22

(C40/C41)

0 to 3

For Display Mode only

A21 FR

(FLW RATE (mA))

R

3.800 to 20.500

mA

3

For Display Mode only

A30 FTL

(TOTAL)

R

0 to 99999999

B30

(D10)

0 to 7

For Display Mode only

A31 RTL

(REV TOTAL)

R

0 to 99999999

B30

(D10)

0 to 7

For Display Mode only

A32 DTL

(DIF TOTAL)

R

-99999999

to 99999999

B30

(D10)

0 to 7

For Display Mode only

A60 —

(SELF CHECK)

R

Good

Error

See “6.5 Alarm Functions”.

(2) Item B (Menu B): Easy Setup items
Those parameters with a high frequency of use have been grouped together in Menu B. All basic functions can

be controlled using only the parameters from this block.

Item

Name

R/W

Data range

Units

Position

of decimal

point

Default value

(*): Indicated item

Description

Display unit

(BRAIN)

Display unit

/BRAIN

B00 Easy Setup

(EASY SETUP)

B10 Language

(LANGUAGE)

W

English

Japanese

French

German

Italian

Spanish

English

Selects the language used for

the display unit. Linked with

H30.

B20 Flow Damping

(FLOW DAMPING)

W

1.0 to 200.0

s

1

5.0 s

Sets damping time. Linked with

C11.

B21 Base Flow Unit

(FLOW UNIT)

W

Ml(Megaliter)

m

3

kl (Kiloliter)

l (Liter)

cm

3

m

t

kg

g

kcf

cf

mcf

Mgal (US)

kgal (US)

gal (US)

mgal (US)

kbbl (US Oil)

bbl (US Oil)

mbbl (US Oil)

ubbl (US Oil)

kbbl (US Beer)

bbl (US Beer)

mbbl (US Beer)

ubbl (US Beer)

ft

klb (US)

lb (US)

m (*)

Selects flow units for the flow

rate span. Linked with C40.