NORMAL OPERATING CONDITIONS

Ambient Temperature:

General-purpose Use:

–40 to +55°C (–40 to +131°F)

Explosion proof type:

In the case of explosion proof type, refer to de-

scription of “ Ambient Temperature” “Enclosure” in

“HAZARDOUS AREA CLASSIFICATION”

*1: Minimum temperature should also be limited according

to minimum fluid temperature of flow tube’s specifica-

tion.

Refer to description of “Fluid Temperature and Pressure”.

*2: Indicator operating range: –20 to +55°C (–4 to +131°F)

Ambient Humidity: 0 to 100%

Lengthy continuous operation at 95% or more is not

recommended.

Fluid Conductivity:

Size 25 to 200 mm (1 to 8 in.): 10 μS/cm or larger

Note: For a fluid with large flow noise (pure water, pure

alcohol or others), or a fluid with low conductivity and

low viscosity, the output fluctuates and is impossible

to measure accurately.

Output Fluctuation:

The output fluctuates depending on the fluid condi-

tions and damping settings.

The following table shows the output fluctuation as

a rough guideline at flow velocity near 100% of flow

span (damping: 5 s)

• Size 25 to 100 mm (1 to 4 in.)

Fluid

Conductivity

[μS/cm]

Fluctuation (% of rate) as a rough guideline

Flow Span

2.0 m/s

Flow Span

4.0 m/s

10

3.0% or less

7.0% or less

50

1.0% or less

1.0% or less

100

0.5% or less

0.5% or less

500

0.5% or less

0.5% or less

• Size 150 to 200 mm (6 to 8 in.)

Fluid

Conductivity

[μS/cm]

Fluctuation (% of rate) as a rough guideline

Flow Span

2.0 m/s

Flow Span

4.0 m/s

10

5.0% or less

Non-recommendation

50

2.0% or less

3.0% or less

100

1.0% or less

1.0% or less

500

0.6% or less

1.0% or less

Measurable Flow Rate Range:

SI Units (Size: mm, Flow rate: m

3

/h)

Size

(mm)

0 to Min. Span Flow

Rate (0.3 m/s)

0 to Max. Span Flow

Rate (10 m/s)

25

0 to 0.5302 m

3

/h

0 to 17.671 m

3

/h

40

0 to 1.3572

0 to 45.23

50

0 to 2.1206

0 to 70.68

65

0 to 3.584

0 to 119.45

80

0 to 5.429

0 to 180.95

100

0 to 8.483

0 to 282.74

150

0 to 19.090

0 to 636.1

200

0 to 33.930

0 to 1,130.9

English Units (Size: in., Flow rate: GPM)

Size

(in.)

0 to Min. Span Flow

Rate (1 ft/s)

0 to Max. Span Flow

Rate (33 ft/s)

1.0

0 to 2.335 GPM

0 to 77.80 GPM

1.5

0 to 5.253

0 to 175.0

2.0

0 to 9.337

0 to 311.2

2.5

0 to 14.59

0 to 486.2

3.0

0 to 21.01

0 to 700.2

4.0

0 to 37.35

0 to 1244

6.0

0 to 84.03

0 to 2800

8.0

0 to 149.4

0 to 4979

Fluid Temperature and Pressure:

The following figure shows maximum allowable fluid

pressure for the flowtube. Further fluid pressure

should also be limited according to flange rating.

F1206.ai

25 to 50 mm (1.0 to 2.0 in.)

(flange type, wafer type)

65 to 200 mm (2.5 to 8.0 in.)

(flange type, wafer type)

Pressure

(MPa)

4 (580)

2 (290)
1 (145)

−0.1 (-14.5)

−40

(-40)

0

(32)

40

(104)

100

(212)

Temperature °C (°F)

−10

(14)

130

(266)

*1

*1: For wafer types of 40 to 200 mm (1.5 to 8.0 in.), and for

carbon steel flange types (process connection code:

C**) of 150 to 200 mm (6.0 to 8.0 in.), the minimum fluid

temperature is –10°C (+14°F).

*2: For fluid temperature of the explosion proof type, refer

to descriptions of “HAZARDOUS AREA CLASSIFICA-

TION”.

Vibration Conditions:

9.8 m/s

2

or less (frequency of 500 Hz or less)

Note: · Level of vibration is in conformity with IEC 60068-2-6

(SAMA 31.1-1980).

· Avoid locations with much vibration where the pipe

vibration frequency is 500 Hz or more. Such a condi-

tion may cause damage to the instrument.