<6. PARAMETER DESCRIPTION>

6-1

IM 01E30D01-01EN

6. PARAMETER DESCRIPTION

6.1 Parameters

With the exception of parameters that were

specified by the customer upon ordering, all

of the internal parameters will initially be set to

default values. Actions such as the modification of

display details can then be carried out whenever

necessary.

IMPORTANT

Make sure to keep the instrument’s power

on at least for 30 seconds after you set the

parameters. If you turn the power off immediately

after the parameters are set, the settings will be

canceled.

NOTE

In order to ensure that correct flow rate data can

be acquired, it is crucial that the nominal size,

flow rate span, and meter factor of the combined

flowtube are set. When the AXR integral

flowmeter is ordered, the nominal size and

meter factor will be set upon shipment from the

manufacturing plant, and these will not require

additional setting.

If a flow rate span was specified upon ordering,

this will be set before shipment. If this is not

the case, however, it will be necessary for the

appropriate value to be set by the user.

6.2 Parameter Lists

Parameter lists are comprised of the following items.

Item Name

Data

range

Default

value

Units Position of

decimal point

R/W

Description

T0601.ai

Describes the parameter content.

The units corresponding to the data range.

R: Read only

W: Writable

Parameter number.

The default value (as setup upon plant shipment).

If marked “(*)”, it indicates that the appropriate

setting has been made in accordance with

ordering information and data for a remote

flowtube to be combined.

This item defines the range of data to be set for

numeric parameters and also defines selection of

data for selection-type parameters.

The position of the decimal point within the data

range and also the allowed relocation range in

terms of places of decimals.

The name of the parameter. Entries in parentheses

indicate the parameter name displayed on the

handheld BRAIN Terminal (BT200).