<11. MAINTENANCE>

11-6

IM 01E30D01-01EN

11.4 Adhesion diagnostic

function

The long-term measurement of adhesive fluids

may adversely affect the flow measurement due

to insulating materials adhering to the electrode,

which could cause the output vibration to increase

or make it impossible to measure the flow rate.

This function checks the level of adhesion by

measuring the resistance value.

NOTE

Adhesion Diagnostic Function

• This function diagnose adhesion using

electrode resistance values.

• The result of adhesion diagnosis is displayed

at

K12:Adh Measure Value.

• When “Adhesion check” has been set for

B41/H11: Display Select 2 or B42/H12:

Display Select 3, the diagnose adhesion

is indicated on the display unit using four

different levels.

• If the judgment value for Level 3 is

exceeded, a warning is displayed; and if

the value for Level 4 is exceeded, an alarm

or warning is displayed depending on the

setting condition at

G29.

• Available conductivity for this function is

limited to:

Nominal size 25 mm or larger: 10 µS/cm

Make sure to use the adhesion diagnostic

function with the greater conductivity than

the above mentioned value.

IMPORTANT

• Parameters cannot be set during execution

(approximately 5 minutes).

• Adhesion diagnosis should only be carried

out when the fluid velocity is completely zero

by closing the valve.

• Flow measurement is not performed during

execution. While adhesion diagnosis is being

carried out, normally a current of 4 mA is

output. When adhesion diagnosis is carried

out, change the control loop to the manual

mode first.

• Adhesion diagnosis cannot be carried out

while an alarm is occurring.