Yokogawa ADMAG AXR User Manual

Page 53

Advertising
background image

<6. PARAMETER DESCRIPTION>

6-5

IM 01E30D01-01EN

Item

Name

R/W

Data range

Units

Position

of decimal

point

Default value

(*): Indicated item

Description

Display unit

(BRAIN)

Display unit

/BRAIN

C43 Flow Decimal Pnt

(FLOW DECIMAL)

W

Auto

0

1

2

3

Auto (*)

Selects decimal point

position for the display unit’s

instantaneous flow rate. Linked

with B24.

C44 Velocity Check

(VELOCITY CHK)

R

0.000 to 99.999

m/s

3

Display of the span setting

using flow velocity (m/s).

C45 Density Unit

(DENSITY UNIT)

W

kg/m

3

lb/gal

lb/cf

kg/m

3

Sets units for density when

mass flow rate is selected.

C46 Mass Flow Density

(MASS DENSITY)

W

0 to 32000

C45

0 to 4

0

Sets density when mass flow

rate is selected (with units from

C45).

C50 User Span Select

(USER SPN SEL)

W

No

Yes

No

Selects whether or not special

units are used for flow rate

units.

C51 Flow User Unit

(FL USER UNIT)

W

8 alphanumeric

characters

Sets the special flow rate units.

C52 Flow User Span

(FL USER SPAN)

W

0.00001 to 32000

C51

0 to 4

100

Sets span when using special

flow rate units.

C60 —

(SELF CHECK)

R

Good

Error

See “6.5 Alarm Functions”.