2 technical data, Technical data, Instrument components and specifications – Leica Biosystems CV5030 User Manual

Page 14

Advertising
background image

14

Instructions for Use V 3.0 RevE – 04/2013

3.

Instrument Components and Specifications

3.2 Technical data

Nominal supply voltages:

100 - 240 V AC, ± 10 %

Nominal frequency:

50 - 60 Hz

Approvals

CE, CSA C/US

Fuse:

Automatic fuseT5A

Nominal power:

100 VA

Dimensions (W x H x D):

Hood closed: 420 x 600 x 550 mm

Hood opened: 420 x 980 x 550 mm

Width (from the left to the right base):

370 mm

Depth (from the posterior base to the front base): 525 mm

Multistainer workstation:

Hood opened: 1620 x 980 x 760 mm

Weight (without packaging material):

approx. 57 kg

Net weight (including packaging material):

approx. 104 kg

Operating temperature range:

+15 °C - +35 °C

Storage temperature range:

–29 °C - +50 °C

Transport temperature range:

–29 °C - +50 °C

Relative humidity:

Maximum 85 % (non-condensing)

IEC 1010 classification:

Protection class 1

Pollution degree 2

Overvoltage installation category II

International protection class:

IP20

Operating elevation:

Up to a max. of 2000 m above sea level

A-weighted noise level:

<65 dB(A), measured in 1 m target range

Sockets

RS 232C, service interface

Uninterruptible power supply (UPS):

The uninterruptible power supply (UPS) should be

designed for a capacity of at least 200 VA over a

time frame of 5 minutes.

Heat dissipation:

100 J/s

Exhaust air:

Hose material

EVA (ethylene vinyl acetate)

Hose length:

3000 mm

Hose diameter:

32 mm

Hose circumference:

41 mm

Exhaust air quantity:

38.5 m

3

/h

Extraction:

Active carbon filter and extraction hose for con-

necting to an external extraction device

Advertising