9 connecting the power supply, 10 installing the accessories, Connecting the power supply – Leica Biosystems CV5030 User Manual

Page 32: Instrument setup

Advertising
background image

32

Instructions for Use V 3.0 RevE – 04/2013

4.

Instrument Setup

4.9 Connecting the power supply

The unit is dimensioned for:

100 to 240 V AC voltage at 50/60 Hz.

Connecting to power supply
• Make sure that the cov-

erslipping machine is

switched off:

The power switch (84) on

the front side must be in

the "0"= OFF position.

The electrical connection is on the left rear side

of the instrument (

Fig. 23

).

• Insert a suitable power cable into the power

input socket (58).

• The serial interface (59) is intended for con-

necting to a Leica ST5020 (multistainer) if both

instruments are being operated together as a

workstation (

refer to Chapter 6

).

58

59

Fig. 23

Cover slip catch tray

The cover slip catch tray (60) is used to hold

cover slips broken by the cover slip mount.

• Hook the cover slip catch tray into the slide

holder outfeed, inserting it towards the right

as shown in

Fig. 24

(52); always ensure that it

is securely in place.

Fig. 24

4.10 Installing the accessories

60

52

84

The instrument MUST be connected to a grounded power socket.

Only use a suitable cable that is appropriate for the local power supply (plug must fit on-site

wall outlet).