Leica CV5030 – Robotic Coverslipper

4. Instrument Setup

• After filling the mountant bottle, check the bottle

neck and threading for cover slip mountant resi-

due and clean it if necessary.

• To achieve optimal coverslipping quality, leave

the filled mountant bottle under a fume hood for

6 - 12 hours covered by a loosely positioned, blue

bottle lid. Only insert the bottle into the instru-

ment after doing this.

• Screw the red lid for the dispenser group tightly

onto the mountant bottle. While doing so, ensure

that the black O-ring is present and positioned

correctly on the dispensing group and the moun-

tant bottle's blue shutter ring.

4.11 Refilling consumables (continued)

Only glass bottles provided by Leica

may be used (

see opt. accessories

and Standard delivery—packing list

).

Other glass or plastic bottles are not

suitable and must not be used on the

instrument.