Instrument setup – Leica Biosystems CV5030 User Manual
Page 35
35
Leica CV5030 – Robotic Coverslipper
4. Instrument Setup
• After filling the mountant bottle, check the bottle
neck and threading for cover slip mountant resi-
due and clean it if necessary.
• To achieve optimal coverslipping quality, leave
the filled mountant bottle under a fume hood for
6 - 12 hours covered by a loosely positioned, blue
bottle lid. Only insert the bottle into the instru-
ment after doing this.
• Screw the red lid for the dispenser group tightly
onto the mountant bottle. While doing so, ensure
that the black O-ring is present and positioned
correctly on the dispensing group and the moun-
tant bottle's blue shutter ring.
4.11 Refilling consumables (continued)
Only glass bottles provided by Leica
may be used (
see opt. accessories
and Standard delivery—packing list
).
Other glass or plastic bottles are not
suitable and must not be used on the
instrument.